Logo

Manufacturer/company:

BAUER KOMPRESSOREN GmbH

Stäblistrasse 8
München
Germany

Number of projects
2
Number of test pieces
3

Documents

BA 2603-1737_Urkunde_01.pdf
Valid until 26.06.2031
BA 2603-1737_Bescheinigung_01.pdf
Valid until 26.06.2031
BA 2603-1737_statement_01.pdf
Valid until 26.06.2031
BA 2603-1737_certificate_01.pdf
Valid until 26.06.2031

Test piece

Name of test piece/s UVC reaction tube B-VIRUS FREE-459
Description Model number: B-VIRUS FREE S
Project number BA 2603-1737
Created on 08/28/2026
Valid until 06/26/2031
Test type UVC radiation intensity
Classfication
UVC radiation intensity

UVC radiation intensity (DIN EN 14255-1)

Inactivation of influenza

Rate

[%]

Required dose [mJ/cm2]

Irradiance

[mJ/cm2*s]

Veff

[l]

tLimit

[s]

Fmax

[m3/h]

90

3.5

50

2.6

0.07

134

99

10.5

0.21

45

99.9

31.5

0.63

15


The following tables show the values which were obtained according to literature data from Bianco et. al., 2020[1] for the inactivation of SARS-CoV-2 as a human-pathogenic virus:

TP01:

UVC radiation intensity (according to Bianco et. al., 2020)

Inactivation of SARS-CoV-2

Rate

[%]

Required dose [mJ/cm2]

Irradiance

[mJ/cm2*s]

Veff

[l]

tLimit

[s]

Fmax

[m3/h]

99.9

3.7

50

2.6

0.07

127

total

kill

16.9

0.34

28



[1]    Bianco, Andrea; Biasin, Mara; Pareschi, Giovanni; Cavalleri, Adalberto; Cavatorta, Claudia; Fenizia, Claudio et al. (2020): UV-C irradiation is highly effective in inactivating and inhibiting SARS-CoV-2 replication. In: medRxiv. DOI: 10.1101/2020.06.05.20123463